<mains_artefact>

Intra-operative MEP Monitoring: Case Example


Initial peripheral EMG responses were contaminated with severe 50Hz artefact.

This was traced to a syringe pump connected to mains which had been inadvertently fixed to the same pole as the patient interface box.

With the syringe pump removed, recordings were much improved (lower trace).

Back to Spinal Monitoring

^TOP FEEDBACK | SITE INDEX | DISCLAIMER | CONTACT | COPYRIGHT | PRIVACY
Printer-friendly version

Last Modified: 27-07-2004 14:26:54